Pb/U fractionation during Nd : YAG 213 nm and 266 nm laser ablation sampling with inductively coupled plasma mass spectrometry
| Date Published |
10/2000
|
|---|---|
| Publication Type | Journal Article
|
| Authors | |
|---|---|
| LBL Report Number |
LBNL-46626
|
| Abstract |
Elemental fractionation during laser ablation sampling was investigated by measuring Pb/U ratios in NIST 610 synthetic glass. Two Nd:YAG lasers with wavelengths of 213 and 266 nm were used to ablate the sample into an inductively coupled plasma mass spectrometer. Pb/U fractionation was observed to be similar for both laser wavelengths, and dependent on the irradiance. For representative Pb/U measurements, the necessary laser irradiance should be >0.6 GW/cm2. However, if the laser beam is initially focused close to the sample surface, fractionation increases and is influenced by the formation of a crater during repetitive pulsing at a single sample location. As the ratio of crater depth to radius increases, plasma sampling and/or an effective irradiance decrease could cause additional fractionation. A good correlation was found between the fractionation of 14 elements in NIST 610 glass and the logarithms of their oxide melting temperatures. |
| Journal |
Applied Spectroscopy
|
| Volume |
54
|
| Year of Publication |
2000
|
| Issue |
10
|
| Number |
10
|
| Pagination |
1435-1442
|
| Keywords |
|
| Organizations | |
| Research Areas | |
| Download citation |