TY - JOUR KW - Measurements KW - Measurement KW - Solution AU - Richard E Russo AU - Dorys Rojas AU - Piotr B Robouch AU - Robert J Silva AB -

A photoacoustic spectrometer was developed for remote optical absorption measurements in aqueous solutions using an 85‐m optical fiber to deliver pulsed tunable dye laser radiation to a sample cuvette located in a glove box. The spectrometer was tested using aqueous solutions of praseodymium ions. Beer’s law was verified down to a concentration of 8×10−6M for an equivalent absorptance of 3.2×10−5 cm−1.

BT - Review of Scientific Instrumentation C2 - LBNL-29512 DA - 12/1990 DO - 10.1063/1.1141544 IS - 12 LA - eng LB - Laser N2 -

A photoacoustic spectrometer was developed for remote optical absorption measurements in aqueous solutions using an 85‐m optical fiber to deliver pulsed tunable dye laser radiation to a sample cuvette located in a glove box. The spectrometer was tested using aqueous solutions of praseodymium ions. Beer’s law was verified down to a concentration of 8×10−6M for an equivalent absorptance of 3.2×10−5 cm−1.

PB - AIP Publishing PY - 1990 ST - Rev. Sci. Instrum. T2 - Review of Scientific Instrumentation TI - Remote photoacoustic measurements in aqueous solutions using an optical fiber VL - 61 ER -